Main menu



MPIII Temperature Data Logger
Operating Temperatures
MPIII Temp : -
MPIII HiTemp : * +100°C to +360°C with Thermal Pack
MPIII HiTemp EXT : * -
*MPIII HiTemp & MPIII HiTemp EXT Tracers require a Thermal Pack for use above +140°C
Temperature Accuracy :
MPIII Temp ± 0.1°C
MPIII HiTemp : ± 0.5°C
MPIII HiTemp EXT : ± 0.25°C at < 200°C / ± 0.50°C at > 200°C
Operating Pressure (For All Three Models) : Up to 10 BAR
Size : Width: 18mm x Length: 20mm) **
Weight : 15 grams **
Material : 316 stainless steel and sapphire IR windows
Sensor Elements : Temperature : Thermistor or RTD
**Size and weight are shown for Tracer body only.
Memory : 16,000 time/temperature observations
Sample Rates Operator programmable (1 second to 18 hour intervals)
Calibration :
Factory calibrated (NIST traceable)
Field calibration (MPIII Temp only)
Battery Type : Lithium (field-
Clock Accuracy : 26 seconds per 24 hours
Certifications

MPIII Relative Humidity Data Logger
Operating Temperature : +10°C to +85°C
Operating Humidity : 0% to 100% RH, non-
Temperature Accuracy : ± 0.1°C
Humidity Accuracy : ±2 %RH at 25°C and 55°C (Non-
Size : Width: 18mm x Length: 66mm
Weight : 48 grams
Material : 316 stainless steel and sapphire IR windows
Sensor Elements
Temperature : Thermistor
Humidity : Capacitive polymer (field-
Memory :
8,000 time/temperature observations
8,000 time/humidity observations
Sample Rates : Operator programmable (1 second to 18 hour intervals)
Calibration RH :
Field calibration to ± 3% RH accuracy
Factory calibrated (NIST traceable)
Battery Type : Lithium (field-
Clock Accuracy : 26 seconds per 24 hours
Certification : LCIE EEx (for intrinsic safety), EC

MPIII Pressure Data Logger
Operating Temperature : +25°C to +140°C
Operating Pressure : 0 PSIA (0 BAR) to 72.5 PSIA (5 BAR)
Temperature Accuracy : ± 0.25°C
Pressure Accuracy : ± 0.45 PSIA (± 0.03 BAR)
Size : Width: 18mm x Length: 48mm
Weight : 55 grams
Material : 316 stainless steel and sapphire IR windows
Sensor Elements
Temperature : Thermistor
Pressure : Strain gauge bridge
Memory
8,000 time/temperature observations
8,000 time/pressure observations
Sample Rates : Operator programmable (1 second to 18 hour intervals)
Calibration : Factory calibrated (NIST traceable)
Battery Type : Lithium (field-
Clock Accuracy : 26 seconds per 24 hours
Certifications : ATEX (for intrinsic safety) -

MPIII Dual Probe Temperature
Dual Probe Temperature Logger
Operating Temperature : -
Temperature Accuracy ± 0.1°C
Operating Pressure : Up to 10 BAR
Physical Size: Width: 18mm x Length: 20mm
Weight: 15 grams
Material: 316 stainless steel and sapphire IR windows
Memory: 16,000 time/temperature observations
Sample Rates: Operator programmable (1 second to 18 hour intervals)
Calibration:
Factory calibrated (NIST traceable)
Field calibration (MPIII Temp only)
Battery Type: Lithium (field-
Certifications: LCIE EEx (for intrinsic safety), EC
Probe Type Rigid Flex
Rigid Probe Lengths 1" , 1.5" , 2" , 3" , 4" , 5"
Flex Probe Types Silicone
Stainless Steel
Flex Probe Tips Taper Tip
Needle Tip
Flex Probe Lengths 6" or 12"
Sensor Element Thermistor RTD