Main menu:
Coatings & Industry > Corrosion + Ageing > Continuous condensation humidity chamber
CONTINUOUS CONDENSATION HUMIDITY CHAMBER according ISO 6270
Water set up temperature is 40 °C to match condition imposed by ISO 6270 standard. Standard speaks about one air temperature, measured at 25 mm from sampling, uniform and constant at ±2 °C and not less than 35 °C, meaning a temperature of 37 °C ±2 °C.
Ambient temperature outside of apparatus must be held at 23 °C ±2 °C.
(1) Welded polypropylene tank
(2) PVC cover holding 20 samplings
(3) Digital thermometer giving exposure sampling temperature
(4) Feeding of water with constant level
(5) Drain pipe
(6) Water temperature regulation system
(7) Main electrical wire
Ref. 115-1012M4
Box made of Polypropylene - Size: 110 x 33 x 20 cm
Cover with double slope (angle of 15° to the horizontal) designed for supporting 20 panels of 150 x 100 mm
Handles for easy cover removing
Electrical heating devise located in the bath
Temperature regulator precision is 2 °C at 40°C
Digital Temperature display with reading at 0.1 °C
Digital Control Thermometer located on the cover
Water bath constant level by automatic water feeding .
Power : 1000 W - 220 V -50 Hz
Important : For this test, we advise to use demineralized water (2 bar pressure)